| Specification | Source-listed value |
|---|---|
| Source model: model | ResMed AirSense 11 family: AutoSet (including AutoSet for Her), Elite and CPAP |
| Source model: region | Americas multilingual user guide, English section |
| Source model: document revision | 398101/1, 2021-04 |
| Source model: manufacturer typical power consumption (W) | 56.1 |
| Source model: manufacturer typical power consumption va | 111.5 |
| Source model: manufacturer peak power consumption (W) | 73.2 |
| Source model: manufacturer peak power consumption va | 137.6 |
| Source model: power supply label (W) | 65 |
| Source model: AC input range (V) | 100-240 |
| Source model: AC input frequency (Hz) | 50-60 |
| Source model: AC input current a | 2 |
| Source model: separate aircraft input | 115 V, 400 Hz, 1.5 A; separate aircraft-use row, not a consumption test |
| Source model: DC output (V) | 24 |
| Source model: DC output tolerance (V) | 1 |
| Source model: DC output a | 2.71 |
| Source model: documented components | HumidAir 11 Standard tub; ClimateLineAir 11 heated tubing or SlimLine tubing; 65W AC adaptor. Component list is not the tested power configuration. |
Calculate running cost
Enter electrical input watts from your appliance label or measurement. The reference does not establish a single running wattage.
Hours, days and rate start with the running-cost calculator's example, not your usage or tariff. Replace them with your own values. The result is in US dollars.
The Americas AirSense 11 user guide places the consumption entries under its power supply specifications. It lists 56.1 W typical and 73.2 W peak. The guide covers AutoSet, Elite and CPAP devices. Keep that family and document edition attached to the values. The power row does not identify a tested pressure, humidifier setting, tubing choice or averaging interval. It cannot establish a separate consumption figure for each configuration.
The supply rating describes a different field
The same block names a 65 W power supply unit and lists a DC output of 24 V and 2.71 A, with a stated voltage tolerance. These are supply specifications. They should not replace the guide's separately labeled consumption entries. The published peak is also not described as a startup surge with a duration. This page does not resolve those different labels into an assumed maximum load or a supply recommendation.
Other test tables do not define the power test
The technical pages include separate flow, sound and humidifier performance results. Their test conditions belong to those results. They are not stated as the conditions for the electrical consumption row. In particular, a humidifier output test does not supply an electrical energy reading. Without a matched power test, this reference cannot divide the total between the blower, heated tubing and humidifier or assign watts to a chosen setting.
Keep energy records separate from care decisions
An energy record needs a defined interval and a clear measurement boundary. The typical watt label alone does not establish the energy used in a night, and the peak label does not establish backup runtime. This electrical reference offers no treatment changes, accessory substitutions or battery sizing. Questions about continued operation and approved power arrangements belong with the device maker and care provider. Keep that review separate from an electricity cost calculation.
Sources
- ResMed AirSense 11 User guide, Americas, 398101/1 2021-04. Accessed 2026-10-22.